Do not overtighten the fitting; doing so may cause a leak.
HPVA Series Operator’s Manual
Connecting Gases
Mar 2011
6-9
3. Check for leaks at the high-pressure side of the regulator and in the connector.
HIgh-pressure Gauge
Gas Bottle Shut-
Off valve
Regulator Control
Knob
Regulator
Connector Nut
a. Turn the regulator control knob fully counterclockwise.
b. Slowly open the gas bottle shut-off valve, then close it.
c. Observe the pressure on the high-pressure gauge.
d. If the pressure is stable, proceed with the next step.
If the pressure decreases, tighten the regulator connector nut until it becomes stable.
5. Purge the air from the lines.
Regulator
Shut-Off valve
Gas Bottle
Shut-Off valve
a. Turn the regulator shut-off valve counterclockwise to open.
b. Open the gas bottle shut-off valve to flow gas.
Содержание HPVA Series
Страница 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Страница 4: ......
Страница 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Страница 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Страница 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Страница 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Страница 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Страница 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Страница 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Страница 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...