Connecting Gases
HPVA Series Operator’s Manual
6-10
Mar 2011
c. Close the regulator shut-off valve to stop flow.
d. Close the gas bottle valve.
5. Set the instrument pressure.
Regulator
Control knob
Low-pressure Gauge
Regulator
Shut-Off
valve
Gas Bottle
Shut-Off valve
a. Turn the regulator control knob clockwise until the low-pressure gauge reads 15 psi
(1.03 bar).
b. Open the regulator shut-off valve.
c. Open the gas bottle shut-off valve and flow gas for 10 to 30 seconds.
d. Close the gas bottle shut-off valve.
5. If you disconnected the gas line to the instrument inlet, reconnect it now.
6. Ensure that the gas pressure is set to the following:
Analysis gas
1500 psi (103 bar) for an HPVA-100
3000 psi (206 bar) for an HPVA-200
Helium gas
500 psi (34.4 bar)
Compressed air
75 to 80 psi (5.2 to 5.5 bar)
7. Adjust the gas metering valve for proper gas flow. Refer to the next topic.
Содержание HPVA Series
Страница 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Страница 4: ......
Страница 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Страница 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Страница 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Страница 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Страница 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Страница 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Страница 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Страница 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...