Connecting Gases
HPVA Series Operator’s Manual
6-8
Mar 2011
Connecting a Replacement Gas Bottle
Move the replacement bottle close to the instrument and tether it into place.
A power failure or loss of cryogen can result in dangerous pressures in the sample
tube. The HPVA uses pressure relief valves to vent this pressure into the instrument
cabinet and return the instrument to a safe condition. When using toxic or flammable
gases, additional venting of the cabinet may be required.
When connecting hazardous gases, be sure to vent properly and follow the safety
procedures established for your lab.
1. Use an appropriate cylinder wrench to remove the protective cap from the replacement bottle.
2. Attach the gas regulator to the connector on the gas bottle. Hand-tighten the nut, then use an
appropriate wrench to tighten an additional 3/4 turn.
Содержание HPVA Series
Страница 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Страница 4: ......
Страница 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Страница 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Страница 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Страница 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Страница 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Страница 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Страница 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Страница 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...