HPVA Series Operator’s Manual
Preparing the Sample
Mar 2011
2-5
7. Subtract the
Weight of sample cylinder
(recorded earlier) from the
Weight of sample cylinder plus
sample.
The value obtained is Weight of sample, which you will enter when starting the
experiment.
Assembling the Sample Holder
1. Assemble the sample holder by placing the gasket on the sample cylinder as shown below.
Gasket
Note that the gasket can be used only once. Reusing the gasket may
cause leaks at the sample holder connection.
Two types of gaskets are available. One is filtered, the other one is not.
It is best to use a filtered gasket for fine powders. The non-filtered
gasket is sufficient for other types of materials.
2. Place the upper portion of the sample holder on the cylinder and hand-tighten the connector nut.