HPVA Series Operator’s Manual
System Configuration Window
Mar 2011
5-11
Free Space Tab
Vent Time
The amount of time the vent remains open after the third helium
fill/vent cycle.
Vent Expand
Vent Pressure
During the adsorption experiment, the system will be considered
vented if at or below this pressure.
Expand Time
During the analysis, the system will open valve 1 to dose the
sample for this amount of time before closing the valve and
waiting for equilibration.
Dose
Pressure
The pressure the manifold will dose to prior to the free space
analysis. The recommended value is 10 bar.
Tolerance
The tolerance of the dosing pressure for the free space analysis
( + this value).
Max Equil Var
The system is considered at equilibrium if the pressure does not
vary by more than this pressure during a one minute period.
Содержание HPVA Series
Страница 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Страница 4: ......
Страница 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Страница 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Страница 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Страница 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Страница 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Страница 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Страница 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Страница 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...