System Configuration Window
HPVA Series Operator’s Manual
5-10
Mar 2011
Adsorption Tab
Pressure Tolerance
Threshold
The pressure at which the dose pressure tolerance below (this
value) is different from the pressure tolerance above (this value).
Above Threshold
The dose pressure must be the value entered here for the
system to continue during dosing.
Below Threshold
The dose pressure must be the value entered here for the
system to continue during dosing.
Temperature Tolerance
The sample temperature must be this value to be
considered at the target temperature.
Helium Flush
The system will cycle helium through the manifold and sample
tube three times after the last experiment to allow safe removal of
the sample.
Vent Pressure
After pressurization with helium, the sample tube and manifold
will be vented to this pressure.
He Fill Pressure
The helium fill pressure.
Содержание HPVA Series
Страница 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Страница 4: ......
Страница 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Страница 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Страница 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Страница 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Страница 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Страница 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Страница 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Страница 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...