HPVA Series Operator’s Manual
Measuring Free Space
Mar 2011
B-7
21. Place the strap around the dewar cover and tighten.
22. Insert the temperature probe into the dewar.
23. Click
OK
.
Do not remove the sample tube from the analysis port between the
ambient temperature volume analysis and the analysis temperature
volume analysis. Doing so will cause the free space analysis to be
compromised.
24. When the analysis is complete, the results will be available in the Excel HPVA macro as described
in the next topic.