Trigger output voltage (V) is out of range. Trigger will not be set.
Cause:
While in the selective adsorption dosing mode, the requested trigger signal
voltage was over 10 volts or under -10 volts.
Action:
Verify that the selective adsorption trigger voltage is within the range of
+10 to -10 volts.
Trigger signal is not ready.
Cause:
While in the selective adsorption dosing mode, the trigger signal to the
external analytical instrument (GC or mass spec) could not be enabled.
Action A:
Verify that the selective adsorption dosing mode was properly selected.
Action B:
Verify proper installation of the HPVA software and that no additional
updates from National Instruments have been installed.
Action C:
Verify that the board and external cables are properly connected.
Unable to load UI.
Cause:
The application initialization failed to load due to unavailable resources.
Action:
Verify that there is available free memory on the computer to run the HPVA
software.
Unable to open instrument log <(path-to-log)>
Cause:
The instrument log file failed to open.
Action:
Verify that you have write permissions to the directory specified for your
data files.
Unable to open template file.
Cause:
The selected template file has been corrupted or is not a proper template
file.
Action:
Select another template file for analysis.
Содержание HPVA Series
Страница 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Страница 4: ......
Страница 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Страница 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Страница 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Страница 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Страница 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Страница 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Страница 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Страница 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...