Aicure ANUP5255V2
Setting and UV irradiation by external devices
48
10.2
EXT.IN parallel signal control
UV irradiation can be controlled under parallel signals by connecting the “EXT.IN” terminal on the
back of the AICURE body with an external device.
■
EXT.IN terminal specifications
■
Input signal
Switch operations on the external device side can substitute for switch operations on the front panel
of the body.
When “Control” of the setting mode is set at “Manual”, priority is given to the following manual
controls.
Pin No.
Input signal
+
GND
Operation
Power ON/OFF
(Lamp ON/OFF)
* Operates the same as the “LAMP ON/OFF”
switch does.
①
⑨
When the contacts short, the lamp
starts to illuminate.
When the contacts open, the lamp
goes out.
Intensity check
②
⑩
When the contacts short, intensity
check begins.
When the contacts open, intensity
check terminates.
UV irradiation (Auto, UV irradiation begins)
* Operates the same as the “AUTO” switch
does.
③
⑪
When the contacts short (one shot),
Auto UV irradiation begins.
UV irradiation (Manual, Shutter Open/Close)
* Operates the same as the “MANUAL” switch
does.
④
⑫
When the contacts short, manual UV
irradiation begins (the shutter opens).
When the contacts open, UV
irradiation terminates (the shutter
closes).
When "Control" of the setting mode is set at "MANUAL", an intensity setting can be selected from
“Intensity selection” UV1 to UV4 by combining manual control signals.
Intensity selection (Intensity
control motor operation)
UV irradiation
ON/OFF
Intensity pattern
Auto signal
③─⑪
Intensity
check signal
②─⑩
Manual signal
④─⑫
UV irradiation (Shutter Open/Close)
Short
ON
Pattern UV1
Open
Open
Open
OFF
Short
ON
Pattern UV2
Open
Short
Open
OFF
Short
ON
Pattern UV3
Short
Open
Open
OFF
Short
ON
Pattern UV4
Short
Short
Open
OFF
(Aicure side: D-Sub 15 pins, female)