Aicure ANUP5255V2
UV irradiation
41
9.2
Operation on the main screen
UV irradiation, “UV control” ON/OFF switching, and “Intensity check” should be performed on the
“main screen”.
On “the main screen”, the settings and lamp use status can be checked.
UV control OFF
UV intensity 0800 mW
Manual intensity control 40%
Back
Next
On
The functions of the “
▲
”, “
▼
” and “SET” buttons vary
depending on the screen.
The three buttons should be operated according to the
functions shown in the lower part of each screen.
[Ex.] When the above item selection screen is shown
▲
(Back)
Moves to the option immediately above the
current option.
▼
(Next)
Moves to the option immediately below the
current option.
SET (On)
Turns UV control to ON (in UV control OFF
situations)
When in “the setting mode”, pressing the “
▲
” button and the
“SET” button at the same time terminates the setting mode,
calling up the main screen.
If no buttons are pressed for one minute, the setting mode
automatically terminates, and the main screen appears.