14
Technical data
5200-en_V2.0_06/20
105
14
Technical data
Interfaces
USB, LAN (Ethernet), RS-232, WLAN
Measurable particle sizes
0.18 – 18 µm
3 measuring ranges
Measurable particle mass
0 – 10 mg/m
3
Measured values
PM
1
, PM
2,5
, PM
4
, PM
10
, PM
Total
, C
n
Air pressure, air temperature, relative hu-
midity
Size channels
64 (32 per decade)
Measuring principle
Optical light scattering by single particles
Quantity measuring range C
n
0 - 20,000 particles/cm
3
Time resolution
1 s - 24 h
In type-approved operation: 15 min
Volume flow rate
4.8 l/min
≘
0.3 m
3
/h
Light source
LED
Power consumption
Normal operation: 60 W
Maximum: 200 W
Data storage
5 GB
Operating system
Windows Embedded for Point of Service
(WEPOS)
Electrical connection
115/230 V, 50/60 Hz
Fuse
T2A
Weight
Controller: 9.30 kg
Sampling head: 2.25 kg
Sampling tube: 4.50 kg
Ambient temperature during
operation
Fidas 200 and Fidas 200 E: +5°C to +40 °C
Fidas 200 S: -20°C to +50 °C
Noise emissions
< 70 dB(A)
Sampling system
Aerosol drying with IADS
(Intelligent Aerosol Drying System)
Sampling head
Passive sampler Sigma-2