5
Operation and settings
68
5200-en_V2.0_06/20
Meaning of parameters and entries
Parameter
Entry
Meaning
Type=
Fidas 200
Type designation of device group. Includes the models
Fidas 200 S and Fidas 200 E.
Ser#
XXXX
Serial number of device. Please do not change.
Password=
XXXX
Password for internal use. Please do not change.
User_device=
XXX
User-defined ID (3 characters)
Fidas 100_enabled=
yes
Plug-in for the top level user interface. Please do not
change.
Fidas 200/210_enabled=
yes
Plug-in for the top level user interface. Please do not
change.
Fidas 300/310_enabled=
yes
Plug-in for the top level user interface. Please do not
change.
My-palas.com_au-
tostart=
no
Initializes automatic connection of the device to the
Palas website every time the device is started. Facili-
tates data upload and remote maintenance. Prerequi-
site: The device is registered on the website.
Velocity_calibrated=
9.3 m/s
Particle velocity setpoint from factory calibration
PM10_slope=
1.000
Slope of fraction PM10. Default = 1. The value can be
changed on the basis of the TÜV (German technical in-
spection association) report 936/2122795.
PM10_intercept=
0
Axis intercept of fraction PM10. Default = 0. The value
can be changed on the basis of the TÜV (German tech-
nical inspection association) report 936/2122795.
PM4_slope=
1
Slope of fraction PM4. Default = 1.
PM4_intercept=
0
Axis intercept of fraction PM4. Default = 0.
PM2.5_slope=
1
Slope of fraction PM2.5. Default = 1. The value can be
changed on the basis of the TÜV (German technical in-
spection association) report 936/2122795.
PM2.5_intercept=
0
Axis intercept of fraction PM2.5. Default = 0. The value
can be changed on the basis of the TÜV (German tech-
nical inspection association) report 936/2122795.
PM1_slope=
1
Slope of fraction PM1. Default = 1.
PM1_intercept=
0
Axis intercept of fraction PM1. Default = 0.
PMtotal_slope=
1
Slope of fraction PMtotal. Default = 1.
PMtotal_intercept=
0
Axis intercept of fraction PMtotal. Default = 0.
PM_alternative
yes / no
Yes = The mesage ”alternative PM” value is active un-
der “accessories.”
textfile=
yes / no
Yes = saving data in a text file is activated.
textfile_interval=
60s
Interval for saving data in a text file.
PM_autoadjust=
no
Default. Please do not change.
IADS_modus=
1
Mode of drying section IADS