5
Operation and settings
5200-en_V2.0_06/20
69
Parameter
Entry
Meaning
dusttype=
2
Default. Please do not change.
sensor_selection=
2
Default. Please do not change.
Automated_cleaning=
yes / no
Yes = automatic cleaning is activated.
Alarm_threshold=
99999 µg/m
3
Limit of PM fraction that controls the digital alarm (digital
out)
Alarm_value=
PM10
Specification of PM fraction that controls the digital
alarm (digital out)
Alarm_email_address=
““
E-mail address to which a message is sent in the event
of an alarm.
Password_service
-1
Password for “Expert user mode”
Calibration_IADS_re-
strict=
yes / no
Yes = calibration is not enabled until the setpoint is
reached.
Calibration_tempera-
ture=
50
Setpoint for drying section IADS at which calibration is
enabled.
Channel_deviation=
3.5
Default. Please do not change.
Weatherstation_con-
nected=
yes / no
Yes = the weather station is connected.
Weatherstation_com-
port=
4
COM port of connected weather station.
Weathersta-
tion_scale_T=
1
Entry of possible slope scales for adjusting the weather
station; obtained from comparison to transfer standard.
Weathersta-
tion_scale_p=
1
Weathersta-
tion_scale_h=
1
Weatherstation_off-
set_T=
1
Entry of possible offset factors for adjusting the weather
station; obtained from comparison to transfer standard.
Weatherstation_off-
set_p=
1
Weatherstation_off-
set_h=
1
Weatherstation_equa-
tion=
X_corr = scale *
x + offset
Function applied to calibrate the weather station.
Discmini_connected=
yes / no
Yes = discmini is connected.
Discmini_comport=
81
COM port for discmini.
Discmini_interval=
300 s
Interval of reported discmini files.
Sensor_selection=
2
Default. Please do not change.
PM_interval=
900 s
Interval for sliding average for PM fractions. 900 sec-
onds pursuant to certificate.
IP_UDP_broadcast=
127.0.0.1
UDP address for data transmission.