Parameter
Entry
Meaning
PLC_interface=
0 / 1 / 2 / 3 / 4 /
5
Communication profile selected upon start.
0 = Modbus
1 = Bayern-Hessen
2 = UDP ASCII
3 = UDP single particle, data flow
4 = Modbus with UDP
5 = serial ASCII
Temperature_compensa-
tion=
yes
Temperature monitoring of LED is activated. Please do
not change.
Temperature_slope=
0.15 / 0.17 /
0.19
Temperature coefficient depending on model:
0.15 for Fidas 200 S
0.17 for Fidas 200
0.19 for Fidas 200 E
Velocity_correction=
yes
Default. Please do not change.
Velocity_calibra-
tion_enabled=
no
Default. Please do not change.
Flow_calibration_en-
abled
yes / no
Yes = calibration of the flow rate is possible (sub-item of
“sensor calibration”)
RSBaudRate=
9600
Baud rate for data transmission
BayerHessen_DA_com-
mand=
60>60,61>61,62
>62,63>63,64>6
4,65>65
Overlay of the address for the communication profile
“Bayern-Hessen”
BayerHessen_history
no
Default. Please do not change.