Verifying and calibrating measuring system
88
5200-en_V2.0_06/20
Calibrating particle velocity
1.
Verify that the adapter hose is placed correctly in the container of
Mon-
oDust 1500
.
2.
Use the blue cursor keys to move the cursor to the right signal peak
below the graph.
ð
The values of the parameters
velocity at cursor (dust)
and
preset velocity
have to be the same.
Maximum deviation permitted:
0.5
m/s.
If the values are the same, calibration of the particle velocity is completed.
If the values are not the same, the value
preset velocity
has to be ad-
justed:
1.
Touch
calibrate velocity
.
ð
A keypad opens.
2.
Enter the value of the parameter
velocity at cursor (dust)
.
3.
Touch
accept
.
ð
Both parameters show the same value.
Calibration of the particle velocity is completed.