Fuse
replacement
If an open fuse is suspected, replace the fuse as follows. Usually
a fuse failure indicates a problem requiring factory repair.
1. Remove power and wait five minutes to allow the charge on
the capacitors to dissipate.
WARNING
Do not operate the unit with the cover removed. Contact with
energized components causes severe shock or burn.
2. Remove the 4 phillips-head screws located on the sides of the
5645.
3. Remove the cover.
4. Check the fuses, F1 and F2, located by the J1 120 V ac
connector. Replace open fuses as follows:
Replace fuse...
With Buss
â type...
F1
MDA 20 amp
F2
MDL 1/2 amp
5. Replace the cover.
3 - 6
5645 Installation and Hardware Manual