8. Press the
TIP SET
Key to store the °F/°C default and
Automatic Temperature Setback time value in system
memory. The Digital Readout will revert to “CAL” and
only the
CH 1
LED will remain lit.
AUTOMATIC POWER DOWN
9. The Automatic Power Down feature is enabled when any time period (10 - 90 minutes) for
Automatic Temperature Setback is entered in step #7. If no time period is selected (i.e., "S - -"
appears on the Digital Readout), the Automatic Power Down feature is disabled. No additional
steps are necessary. The Automatic Power Down feature operates when (and only when) all
channels are in Temperature Setback. For example, power to all channels is turned off 90
minutes after the last Active Channel’s Tip Temperature is set back.
CHANNEL SECTION
10. The
CH 1
LED is now on signifying that Channel 1 is ready to set temperature limits. Perform
steps 11 through 17 to calibrate. Change channels as directed and repeat these steps for each
channel.
Figure 83. Store Temperature Default