49
26. Using the
CH SELECT
Key, select each Active
Channel in sequence, making it the Current Channel
(temperature information displayed on Digital
Readout). Using the procedures described in
previous steps #10 thru 18 and steps #24 thru 25 as
a reference, enter and store desired
SET TIP
Temperatures and appropriate TIP OFFSET
CONSTANTs into system memory. Refer to the “Tip
& Temperature Selection System” charts sent with
your unit and the “Tip & Temperature Selection”
portion of this manual for more detailed information
on selection of the proper Tip, Handpiece and
Temperature Options for your particular application.
NOTE
The PRC 2000 systems incorporate a “Dynamic Offset” feature which
automatically adjusts the Tip Temperature Offset (based on the entered TIP
OFFSET CONSTANT) for any Set Tip Temperature established by the
operator. This ensures the maintenance of true, accurate Tip Temperatures.
Simply stated, any operating Tip Temperature displayed on the Digital
Readout will be correct.
Always set the appropriate TIP OFFSET CONSTANT for the selected
Handpiece/Tip combination (listed in the shaded area on the Tip &
Temperature Selection System Charts) before entering the desired Set Tip
Temperature. The Set Tip Tempe the Dynamically Adjusted Tip
Temperature Offset value (usually different from the entered Tip Offset
Constant) cannot exceed 489°C (912°F). If this limit is exceeded, the
system will automatically lower the maximum possible Set (and Operating)
Tip Temperature accordingly.
Figure 48. Change Current Channel
Содержание PRC 2000
Страница 1: ...1 SYSTEM OPERATION MAINTENANCE MANUAL PRC 2000 SYSTEMS ...
Страница 2: ...2 ...
Страница 3: ...3 MANUAL NO 5050 0313 REV G ...
Страница 17: ...17 Figure 6 Thermal Management Center ...
Страница 23: ...23 Figure 12 Pik And Paste Section ...
Страница 74: ...74 ...
Страница 95: ...95 Figure 122 Pulse Plate Set Up ...
Страница 96: ...96 ...
Страница 102: ...102 ...
Страница 111: ...111 ...
Страница 119: ...119 ...