Figure 5-71 CURS MEAS Menu
The description of the
cursor measurement menu
is shown as the following table:
Function
Menu
Setting
Description
Type
OFF
Voltage
Time
Switch off the cursor measurement.
Display the voltage measurement cursor and menu.
Display the time measurement cursor and menu.
Source
CH1
CH2
Display the channel generating the waveform to
which the cursor measurement will be applied.
When carrying out the cursor measurement, the position of Cursor 1 can be adjusted
with the
VERTICAL POSITION
knob of Channel 1, and that of Cursor 2 can be
adjusted with the
VERTICAL POSITION
knob
of Channel 2.
Perform the following operation steps for the voltage cursor measurement of the channel
CH1:
1.
Press
Cursor
and recall the Cursor Measure menu.
2.
Press the
H2
button and choose
CH1
for
Source
.
3.
Press the
H1
button, the
Type
menu will display at the right of the screen. Then
press the
F2
button to choose
Voltage
for Type, with two purple dotted lines
displayed along the horizontal direction of the screen, which indicating Cursor1
and Cursor2.
4.
According to the measured waveform, adjust the positions of Cursor1 and
Cursor2 by turning the
VERTICAL POSITION
knob
of CH1 and CH2. Cursor
increment window at the left bottom of waveform shows absolute value of
D-value of cursor 1 and cursor 2 and the present position of the two cursors (see
).
Figure 5-72 Waveform of Voltage Cursor Measurement