5.Advanced User Guidebook
72
Figure 5-70
Rise Time:
Time that the leading edge of the first pulse in the waveform takes to rise
from 10% to 90% of its amplitude.
Fall Time:
Time that the falling edge of the first pulse in the waveform takes to fall
from 90% to 10% of its amplitude.
+D width:
The width of the first positive pulse in 50% amplitude points.
-D width:
The width of the first negative pulse in the 50% amplitude points.
+Duty:
+Duty Cycle, defined as +Width/Period.
-Duty:
-Duty Cycle, defined as -Width/Period.
Delay
1→2 :
The delay between the two channels at the rising edge.
Delay 1→2 :
The delay between the two channels at the falling edge.
Duty cycle:
Defines as (the width of the positive pulse)/(Entire period)
The phase difference of two channels:
Phase
: Compare the rising edge of CH1 and CH2, calculate phase difference of two
channels.
Phase difference =
[ (Delay between channels at the rising edge)*PI]/Period.
How to Measure with Cursors
Press the
Cursor
button to display the cursor measurement function menu (
CURS
MEAS
) in the screen.
The Cursor Measurement for normal mode:
The cursor measurement includes
Voltage Measurement
and
Time
Measurement
at