![Oval MAG-OVAL III MGS Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/oval/mag-oval-iii-mgs/mag-oval-iii-mgs_instructions-manual_1662164005.webp)
C
−
102-1-E
5
2. BEFORE YOU BEGIN
Although this flowmeter is built and adjusted exactly to meet the customer specifications and, it is ready to
operate in your service by turning on the power, it is desirable for the operator to read this instruction manual
carefully for the best results in maintaining the stated accuracy and long life before initiating service operation.
You can avoid potential downtime of the equipment by observing the following instructions in particular.
(1) Confirm the operating conditions
The operating conditions and major specifications of this flowmeter appear on the nameplate attached to the
transmitter. Verify that the conditions indicated agree with your particular operating conditions. If you are to
make measurement of corrosive liquids in particular, make sure of the compatibility of electrode materials, etc.
(2) About grounding
An electromagnetic flowmeter requires earth grounding. See the Grounding Instruction on page 12 for proper
grounding.
(3) Use with the flow tube filled with liquid
An electromagnetic flowmeter requires its internal space to be completely filled with the process liquid.
Ensure of a fully filled flow tube during operation.
(4) About field wiring
For intercabling between the electromagnetic flowmeter and receiving instrument, see the section "Wiring
Connections".
Please remember to check terminal numbers for correct connections before operation.
①
②
③
④
⑤
⑥
⑦
⑧
⑨
⑩
⑪
⑫
⑬
⑭
Magnetic Flowmeter
OVAL Corporation
MADE IN JAPAN
MODEL
SERIAL No.
TAG No.
FLUID
FLOW RANGE
MAX. PRESS
MAX. TEMP
ELECTRODE
POWER
DATE
ANALOG OUTPUT
PULSE OUTPUT
CAL. ZERO
CAL. FACTOR
A
Side view from Arrow A
〈
Explanation of items shown on nameplate
〉
No.
Item
Description
(1) Product code
Ex. MGW1-0501T17E-10A
(2) Product No.
——
(3) Instrument No.
As designated
(4) Fluid measured
Ex. Water
(5) Flow range
Ex. 0 to 100m
3
/h
(6) Max. pressure
Allowable max. pressure of flowmeter
(7) Max. temperature
Allowable max. temperature of flowmeter
(8) Electrode material
Ex. SUS316L (depending on selected type)
(9) Power spec.
Ex. 85 to 264VAC 50/60Hz (depending on selected type)
(10) Month/Year of shipment
——
(11) Analog output
4 to 20mA
(12) Pulse output
At factored pulse Ex. 1L/P, At frequency pulse Ex. 1000Hz at F.S.
(13) Calibration zero value
——
(14) Calibration factor value
——