About status display during external operation through communication
When special operation or processing is executed through communication (at the execution of individual
functions, etc.), the following message or code is displayed on the LCD display of the transmitter.
When the operation is stopped or cleared, the message will disappear.
Status Display
Status category
Message displayed
on transmitter
Status code displayed
on transmitter
Description
Simulating output
A.Out.Fi
St-60
Simulating output of analog output is in process.
P.Out.Fi
St-61
Simulating output of pulse output is in process.
S.Out.Fi
St-62
Simulating output of status output is in process.
Self-diagnosis
SELF.d.VoL
St-70
Self-diagnosis-circuit voltage diagnosis is in process.
SELF.d.CoI
St-71
Self-diagnosis-exciting coil, exciting circuit diagnosis
is in process.
SELF.d.SiG
St-72
Self-diagnosis-measurement signal diagnosis is in
process.
SELF.d.PiP
St-73
Self-diagnosis-measurement pipe empty detecting
function is in process.
Error is a state where setting is inappropriate or normal flow measurement is not available due to occurrence of
significant problem. Alarm is a state where measurement is executed but output or display is not available
normally. Status is to report the specific processing or operation (self-diagnosis, etc.) is in process through
communication (not affecting the measurement).
20.2 Other Statuses
About measurement and output in individual status
Status
Measurement
status
Display status (*)
Output operation
Instant flow
rate
Total flow
Analog
output
Pulse output Status output
Loop test
In analog simulating output
Normal
measurement
Normal
display
Normal
totalizing
operation
Simulating
output
Normal
measurement
output
Depends on
output
setting
In pulse simulating output
Totalization
stopped
Normal
measurement
output
Simulating
output
In status simulating output
Normal
totalizing
operation
Normal
measurement
output
Simulating
output
Self-
diagnosis
In circuit voltage monitoring
Depends on
output
setting
In coil exciting circuit
abnormality diagnosis
In measurement error signal
diagnosis
In empty detection diagnosis
In other mode display
NOTE: * Even during loop test (simulating output), flow display shows actual flow rate measured.