Overflow error
Abnormal hardware error
Parameter alarm
H.FAiL
F.Prm2
F.Prm3
00-00
01-00
02-00
03-00
04-00
05-00
06-00
07-00
08-00
09-00
10-00
11-00
12-00
13-00
14-00
PAS.wd
08-01
08-02
08-03
dSP.St
F.Prm1
diAG.S
LP.tSt
A.trim
CHECK
A.Out
P.Out
Er.Out
ALArm
10-01
10-02
10-03
Low alarm point
AL.Lo
High/Low alarm hysteresis
AL.HyS
High alarm point
AL.Hi
S.Out
St.SEL
Parameter secondary item
Parameter major item
Prm.AL
Output saturation alarm
08-04
SAt.AL
OF.Err
Measurement error signal alarm
Empty tube detection alarm
Fixed output state
EmP.AL
08-05
08-06
08-07
FOut.S
Self-diagnosis execution state
08-08
diAG.S
SiG.AL
Analog output at error
09-01
Pulse output at error
09-02
Er.PLS
Er.AnA
Content of parameter
Default
Setting range
On (Valid), OFF (Invalid)
On (Valid), OFF (Invalid)
On (Valid), OFF (Invalid)
On (Valid), OFF (Invalid)
On (Valid), OFF (Invalid)
On (Valid), OFF (Invalid)
On (Valid), OFF (Invalid)
On (Valid), OFF (Invalid)
0.0000E0
0.0000E0
0.0000E0
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
Content of parameter
Default
Setting range
ZEro
dn.SCLE
Can select the analog output status at the time of error detection.
* For details, refer to "ANALOG OUTPUT FUNCTION".
Outputs when flow exceeds the preset measurement range.
Outputs when there is an abnormal error in the flowmeter hardware.
Outputs when parameter setting is inappropriate.
Outputs when analog or pulse output exceeds the preset higher or lower limit.
Outputs alarm when measurement signal is abnormal (noise superimpose, etc.)
Outputs alarm when the electrode of probe is not wetted.
Outputs status while pulse and analog fixed outputs are being generated in a
loop test.
Outputs when self-diagnosis is being carried out by key operation or through
communication.
Sets the pulse output mode at the detection of an error.
* For details, refer to "PULSE OUTPUT FUNCTION".
"dn.SCLE"
: Down scale (2.4mA),
"HoLd", "ZEro" : 4mA,
"no.FunC"
: Not assigned,
"uP.SCLE"
: Up scale (21.6mA)
"ZEro": 0Hz, "no.FunC": Not assigned,
"uP.SCLE": Up scale (110% of max.
output), "HoLd"
Content of parameter
Default
Setting range
Sets the High alarm point. When a flow greater than the preset flow occurs, this alarm turns
on. (Operates when status output function is assigned to High/Low alarm point setting.)
* Unit is subject to the setting of instant flow rate display unit.
Sets the Low alarm point. When a flow smaller than the preset flow occurs, this alarm turns
on. (Operates when status output function is assigned to High/Low alarm point setting.)
* Unit is subject to the setting of instant flow rate display unit.
Sets the hysteresis width of High and Low alarm points.
* Unit is subject to the setting of instant flow rate display unit.
Set the value in the range greater
than Low alarm point and smaller
than High measurement limit flow.
* 5 effective exponent
Set the value in the range smaller than
High alarm point and greater than 0.
* 5 effective exponent
* 5 effective exponent
m/s
modE
**.***
**.**
**.**
"SEL"
"Press and hold SEL"
"Press and hold ENT"
Other
Parameter name
Layer code (Major item ― Secondary item ― Third item)
Parameter
display symbol
ENT
Displays instant flow rate
Total flow
Displays % flow rate
Resettable total flow
Displays flow velocity
Setup mode
Set flow parameter 2
Set flow parameter 3
[Top row display] Selects display
[Bottom row display] Displays total flow
Enter password
Set display
Set flow parameter 1
Set self-diagnosis
Loop test
Analog output trim function
Self-check function
Set analog output
Set pulse output
Set error output
Set alarm
Set status output
Select status output object
*****
*****