13.3 Adjusting Analog Output
The current of analog output has been adjusted at the time of factory shipment. If, however, any error is present
in the output value and measurement value depending on your service environment, adjustment of the current
of analog output is available.
Analog output adjustment begins from "A.trim" of Setup mode.
■
Preparation for adjustment
Connect the ammeter to the analog current loop in series.
(Using the current monitoring function of the connected measurement, control instrument, etc. are also
permitted.)
Be careful not to cause unexpected operation of external instrument due to variation of current.
■
Adjustment of 4mA output
Enter the adjustment mode from "trm.04" (layer code 13-01) of "A.trim" submenu in the Setup mode.
Perform adjustment for 20mA from the menu "trm.20" in the same manner.
After setting, "Press and hold SEL" from the Menu to return to "A.trim" display. Then simulative output is
cleared and normal output state is restored. (Simulative output continues in the "A.trim" submenu.)
Adjustment of 4mA output
SEL
: Changes numeral
Press and hold SEL : Cancel
ENT
: Shifts digit
Press and hold ENT : Fix
Press and
hold ENT
Press and
hold ENT
The numeral on the bottom row of LCD blinks
and 4mA is output from the transmitter.
Enter the ammeter reading at this time in the
same way as that of normal parameter setting.
Confirm the entry by "Press and hold ENT".
When the entry is confirmed, 4mA after
adjustment is output.