6.2 Wiring Connection Diagram
Fig. 6.4 Connecting Terminal Layout
GND terminal location
Details of terminal block
Note: Ground terminal "EARTH" shall be used for grounding. Ground terminal "EARTH" of this
flowmeter is internally connected to "GND".
6.3 Power and Output Signal Terminal Identification
Item
Terminal No.
Label
Description
O
ut
p
ut
s
ig
na
l
TB2
1
PULSE +
Pulse output
(Open drain pulse output, or voltage pulse output)
2
PULSE
−
3
Status output
(Open collector output)
4
STATUS
−
5
Unused
6
Unused
7
Unused
8
Unused
9
Analog output
10
ANALOG
−
11
RS485 +
RS-485 communication
12
RS485
−
P
ow
er
TB1
1
LINE
(
+
)
Power (with DC power: + )
2
GND
Ground
3
NEUT
(−)
Power (with DC power: - )
4
EARTH
Ground
Power fuse
Output signal terminal (TB2)
screwless terminal
Internal ground terminal
External ground terminal
Power terminal (TB1)