12. DISPLAY TRANSITION
12.1 Display Transition (1) Setting Display
"SEL"
"Press and hold SEL"
"Press and hold ENT"
Other
ENT
Displays instant flow rate
Total flow
Displays % flow rate
Resettable total flow
Displays flow velocity
Setup mode
Set password
Set password function
Set password
Set flow display
Instant flow rate units
No. of instant flow rate digits
below decimal point
Total flow units
No. of total flow digits below
decimal point
Set backlight
Screen display update cycle
F.dECP
m/s
modE
t.dECP
t.unit
FL.SEt
Parameter third item
Parameter third item
Pw.SEt
Pw.OPt
F.unit
Set flow parameter 2
F.Prm2
Set flow parameter 3
F.Prm3
Parameter major item
[Top row display] Selects display
[Bottom row display] Displays total flow
%
Enter password
00-00
01-00
02-00
03-00
04-00
05-00
06-00
07-00
08-00
09-00
10-00
11-00
12-00
13-00
14-00
PAS.wd
Parameter name
Layer code (Major item ― Secondary item ― Third item)
Parameter
display symbol
01-01
01-02
01-03
01-04
Set display
dSP.St
Set flow parameter 1
F.Prm1
Set self-diagnosis
diAG.S
Loop test
LP.tSt
Analog output trim function
A.trim
Self-check function
CHECK
Set analog output
A.Out
Set pulse output
P.Out
Set error output
Er.Out
Set alarm
ALArm
Set status output
S.Out
Select status output object
St.SEL
Parameter secondary item
bL.SEt
dP.CyC
Pw.FnC
01-01-01
01-02-01
01-02-02
01-02-03
01-02-04
Times of resettable total overflow
rt.oF
01-02-05
Times of total value overflow
t.oF
01-02-06
01-01-02
Description of item
Content of parameter
Default
Setting range
Content of parameter
Default
Setting range
Content of parameter
Default
Setting range
500
100, 300, 500, 1000 (ms)
Auto
Can change the ON/OFF setting of LCD backlight.
0 to 3 digits
* 0 digit: Indication by integer
0 to 3 digits
* 0 digit: Indication by integer
0
0
As
specified
Depends on
measurement
range
Depends on
measurement
range
As
specified
L, m
3
, kg, t, gal
On (Valid), OFF (Invalid)
OFF
0000 to 9999
0000
Allows selecting valid/invalid of password feature. Selecting
"ON" causes
password entry to be required before entering the Setup menu screen, preventing
the parameter from being changed easily.
Allows setting a password. To validate this setting, the password feature shall be
set to
"On" by password feature setting. Note that password change is impossible
if you forget your password.
Display of optionally resettable total flow
Operation: Reset this display by "press and hold ENT" while resettable total is blinking.
Can change the unit of instant flow rate indicated on the bottom row display.
Note: At the setting change of volumetric flow rate
⇔
mass flow rate, the total value is reset.
* In case of mass flow rate display, density shall be set additionally as fixed conversion factor.
Can change the number of digits below decimal point of instant flow rate indicated
on the top row display.
Can change the unit of total flow indicated on the bottom row display. (Common to accumu-
lated total and resettable total)
* In case of mass flow rate display, density shall be set additionally as fixed conversion factor.
Can change the number of digits below decimal point of total flow indicated on the
top row display.
(Common to accumulated total and resettable total)
Number of times the total value indicated on the bottom row display exceeds the
display digits (8 digits) and returns to
"0". Resetting the resettable total resets the
count to zero.
Number of times the total value indicated on the bottom row display exceeds the
display digits (8 digits) and returns to
"
0
"
.
Can change LCD display update cycle (ms). If viewing is difficult due to
display change, set the interval longer, and if quick review is required,
shorten the interval.
Unit of flow rate (Numerator):
L, m
3
, kg,
t, gal
Unit of time (Denominator): s, min, h, d
ON (lights always), OFF (unlit always)
Auto (Lights during operation, turns off
10 sec. after finishing operation)
Only for reference
Only for reference
**.***
*****
*****
**.**
**.**