Section 3 - Installation & Specifications
ORTHO
™
Workstation for ID-MTS
™
Gel Cards Reference Guide
8
Pub. No.: J55700ENna
2013-09-04
Section 3 - Installation & Specifications
Unpacking Procedure
1. Visually inspect the container for damage and ensure it is in an upright position before it is
opened.
2. Remove power cord from box.
3. Remove the instrument from the box and protective bag.
4. Remove the foam block from the rotor area.
5. Visually inspect the instrument for loose, bent, or broken parts.
Report any damage immediately by contacting the OCD Customer Technical Services number
provided by Ortho Clinical Diagnostics or your distributor.
6. Compare the serial number on the rear panel of the instrument with the serial number in the
packing slip.
If the Instrument is Received Damaged
If you have received an instrument with obvious shipping damage, do not connect power to the
instrument. Contact the
OCD Customer Technical Services
number provided by Ortho Clinical
Diagnostics or your distributor for assistance.
After installation, a quality control program should be followed through the life of the ORTHO
™
Workstation to provide assurance of accurate test results. Never use if the equipment does not
operate properly or is damaged.
Specifications
CAUTION: Operator is advised to read all instructions before plugging the
instrument in.
Dimensions:
Height – 220 mm (8.66 inches)
Width – 575 mm (22.64 inches)
Depth – 325 mm (12.80 inches)
Weight:
10.89 kg (24 lb)
Power Requirements:
100 – 240 V~, 50/60 Hz, single phase
(Input auto-senses voltage/current, no
manual selection required)
Power Consumption:
150 VA