Section 4 - Centrifuge
ORTHO
™
Workstation for ID-MTS
™
Gel Cards Reference Guide
12
Pub. No.: J55700ENna
2013-09-04
6. Press the centrifuge Start button. Centrifuge rotor will accelerate up to speed.
The centrifuge rotor speed is monitored for the duration of the centrifuge cycle. If the instrument
cannot reach speed due to a malfunction, an alarm will sound and an error code will be displayed.
An incomplete cycle voids all results. See Section 6 - Troubleshooting and Error Codes.
You can cancel a cycle in progress by pressing the centrifuge Stop button. An incomplete cycle
voids all results.
At the conclusion of the ten-minute spin, the completion alarm will beep three times and the time
display will flash 00:00.
7. Open the lid and remove the cards.
NOTE: Check the centrifuge rotor regularly for signs of damage. The centrifuge must not be used
if there are signs of damage or if the speed or timer is out of specification
.
Centrifuge Card Holder Replacement
1. Open the centrifuge lid.
2. Pinch the center of the damaged card holder and remove it from the centrifuge rotor.
3. Place the new card holder in the centrifuge rotor and pinch the center of the card holder so that
it sits in place in the centrifuge rotor.
4. Ensure that the card holder swings freely.
5. Close the centrifuge lid.
CAUTION: ID-MTS Gel cards placed incorrectly in the centrifuge card holders may
result in incorrect results. Always verify ID-MTS Gel cards are seated firmly in the
card holder.
CAUTION: Incomplete, interrupted or multiple cycles may produce erroneous
results. When in doubt, discard the cards and repeat the test.
WARNING: If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment may be impaired.
WARNING: A pinch hazard exists if the centrifuge lid is opened before the rotor
stops spinning. Allow the rotor to come to a complete stop before opening the lid.