ORTHO
™
Workstation for ID-MTS
™
Gel Cards Reference Guide
Pub. No.: J55700ENna
9
2013-09-04
Section 3 - Installation & Specifications
Site Specifications
Select a location for the ORTHO
™
Workstation that is convenient for laboratory personnel to
perform testing. The surface must be a stable, level, vibration-free surface, away from direct
sunlight, free from dust, solvent and acid vapors and have an electrical outlet close enough to the
instrument to permit power access without an extension cord. Do not expose it to heat or cold
(close proximity to heating or air-conditioning systems). Never subject it to violent shaking or any
other shock or impact.
\
Environmental Requirements
Electrical Requirements
The ORTHO
™
Workstation is equipped with automatic sensing of input voltage within the range of
100 – 240 V~, 50/60 Hz, single phase. Connect the instrument to the power outlet with the
grounded three-prong power cord provided.
Ventilation Requirements
There are no special ventilation requirements for this instrument.
Fuses:
2 required, T4AH250V - 4 Amp, 250V,
ceramic, time delay, 5x20mm.
Cycle Speed:
1032 rpm ± 10 rpm
Speed Indicator:
4 digits
Timing Indicator:
4 digits
Cycle Timing:
10 minutes ± 10 sec.
(including ramp-up time)
Centrifuge Card Capacity:
10 cards
Incubation Temperature:
37°C ± 2°C
Incubator Card Capacity:
20 cards (two sections of 10)
Noise Level Generated:
Maximum 50 dba at 1 meter (during normal
operation)
WARNING: As with any electrical appliance, place the instrument AWAY from any
source of water, such as a laboratory sink. An electrical hazard may exist when
improperly using the ORTHO
™
Workstation near water.
WARNING: Do not block access to the power cord or Stop button during
ORTHO
™
Workstation operation.
Temperature:
15 – 30°C (59 – 86°F)
Relative Humidity:
15 – 85% (non-condensing)