Section 8 - Maintenance
ORTHO
™
Workstation for ID-MTS
™
Gel Cards Reference Guide
20
Pub. No.: J55700ENna
2013-09-04
Cleaning Under the Incubator (As Needed)
1. Turn off the instrument and unplug the power cord.
2. Open the incubator lid.
3. Using a Phillips screwdriver remove the three screws located on the incubator block.
4. Tilt the incubator block to gain access to the area beneath the incubator.
NOTE: Do not pull the incubator block out or remove the heater harness.
5. Clean any potentially contaminated areas of the system with mild detergent and then with 70%
isopropyl alcohol. Do not scrub excessively.
6
.
Once cleaning is complete, place the incubator block down.
7.
Install the three screws.
8.
Close the incubator lid
Replacing Fuses
1. Turn off the instrument.
2. Unplug the power cord.
3. Pull out and flip fuse lid upwards on rear panel to open.
4. Replace the broken fuse with another one of the same type and value. See Section 3 -
Installation & Specifications.
5. Flip down and push in fuse lid to close.
6. Plug in the power cord and turn on the instrument.
If the instrument does not switch on or the new fuse is not functioning, contact the
OCD Customer
Technical Services
number provided by Ortho Clinical Diagnostics or your distributor for
assistance.
WARNING: Do not open the instrument housing and attempt repairs. If the
troubleshooting table suggestions do not correct the problem, contact the
OCD
Customer Technical Services
number provided by Ortho Clinical Diagnostics or
your distributor for assistance.