Section 5 - Incubator
Section 5 - Incubator
Operational Guidelines
The ORTHO
™
Workstation incubator has one heated block with two sections able to hold up to 10
cards each.
NOTE: The incubator must not be used if the timer or temperature is out of specification.
Operators can set each block timer to any time between and including 1 to 99 minutes, with one
minute increments.
Incubator timed countdown is displayed on the incubator time LED display on the front of the
incubator. When incubation is finished, a completion alarm sounds and the display will flash and
count up indicating over-incubation.
NOTE: If the temperature indicator flashes red during incubation, discard all cards in incubation
process. Avoid incubation until the temperature indicator returns to green.
Procedure- Incubate ID-MTS
™
Gel Cards
No special training is required to operate this instrument.
1. Turn the instrument on and ensure that the incubator lid is closed completely.
2. Wait until the temperature of the ORTHO
™
Workstation is stabilized (10 to 15 minutes) and the
incubator temperature status display is green. If the temperature is not within the required limits
the indicator will flash red and the incubator should not be used.
3. Open the lid and load the cards. When loading cards into the incubator block, make sure the
card is seated fully in the slot. Ensure that the incubator lid is closed completely at all times during
incubation except when loading and unloading cards.
4. To select a desired incubation time other than the factory-set default (15 minutes), press the Up
or Down button.
5. Press the Start/Stop button to start incubation time. The instrument displays the remaining time
continuously. When the incubation is completed, a completion alarm will beep three times and will
flash and count up to show over-incubation time. To stop the timer and reset the display, press the
Start/Stop button.
NOTE: Stopping the incubator timer does not stop the incubator block from heating. Heating will
only stop when the instrument is turned off.
CAUTION: The factory-set default incubation time is 15 minutes.
CAUTION: Once incubation time has been set by the operator, and until the system
is powered off, the displayed starting incubation time shall be the previously set
time. Once the system has been powered off and back on, the starting incubation
time shall return to the factory-set default.
BIOHAZARD: Handle all blood and materials in contact with blood as if capable of
transmitting infectious agents. It is recommended that blood and materials in
contact with blood be handled using established good laboratory practices.
Содержание ID-MTS
Страница 1: ...TRANSFUSION MEDICINE J55700ENna ID MTS Gel Cards Reference Guide ...
Страница 4: ...4 Pub No J55700ENna 2013 09 04 ...
Страница 24: ......
Страница 25: ......
Страница 26: ...Ortho Clinical Diagnostics Inc 1001 US Highway 202 Raritan NJ 08869 USA IVD ...