33
4.
OPTIMIZING INSTRUMENT STABILITY
AND SAMPLE MOUNTING
Optimizing the experiment conditions and sample mounting can greatly improve the quality of the
indentation. During an indentation, environmental stability, sample stability and probe suitability
define the quality of the experimental dataset. In case the indentation results show irregularities or
high noise, it is worth considering the factors as described below.
4.1
Environment stability
The Piuma Nanoindenter is engineered to operate as independent from the environment as
possible. Nonetheless it is still possible for the environmental conditions influence and compromise
measurement results.
Temperature
In order to minimize the measurement error due to temperature drift, sudden room or medium
temperature changes should be avoided when measuring. Verify the Piuma is mounted away from
any air-conditioning, heating or ventilation vent
. If present, the Piuma’s heated stage can be used to
measure samples at physiological temperatures. Be aware that the heater can also introduce drift in
the measurement if the sample or medium temperature has not yet reached an equilibrium state.
Vibrations
To minimize the measurement error due to mechanical disturbances, such as ground vibrations
from other labs, other equipment or appliances, ensure the Piuma Nanoindenter is mounted on a
stable bench. Preferably this bench is not shared with other vibrating machines such as vortexes,
pumps, or large fans. Optionally the Piuma Nanoindenter can be mounted on top of a third-party
stabilization platform, such as an air table or spring table.