4-7
ZUV
User’s Manual
Section 4
CONNECTING TO EXTERNAL DEVICES
Section 4
Connection by an I/O Terminal Block
■
Channel 1 irradiation terminal input (batch control)
This input starts/stops UV irradiation of all channels.
• The input time is 40 ms and 120 ms when SWITCH and CONTACT are set,
respectively.
• Irradiation starts only when the ready output of all channels turns ON when this signal
is input.
• Irradiation of all channels stops when irradiation is in progress for any one of the
channels.
■
Channels 1 to 4 irradiation terminal input (individual control)
This input starts/stops the irradiation of each of the channels from 1 to 4.
• The input time is 40 ms and 120 ms when SWITCH and CONTACT are set,
respectively.
• Irradiation starts when the ready output of the target channel turns ON. (Otherwise,
irradiation of the target channel stops).
Control action is as follows when irradiation log data management is disabled.
Note, however, that the input time is 120 ms or more when CONTACT is set.
Setting ready signals output conditions p.3-21
TRG1 ON
CH1 irradiation
CH1 READY
CH2 irradiation
CH2 READY
40 ms/120 ms
40 ms/120 ms
40 ms/120 ms
200 ms
TRG2 ON
CH2 irradiation
CH2 READY
40 ms/120 ms
40 ms/120 ms
TRG1 ON
CH1 irradiation
CH1 READY
40 ms/120 ms
40 ms/120 ms
40 ms/120 ms
200 ms
200 ms
TRG1 ON
CH1 irradiation
CH1 READY
40 ms/120 ms
40 ms/120 ms
Содержание ZUV Series
Страница 1: ...Smart Curing System User s Manual Cat No Z281 E1 02 ZUV Series Mega Power Type...
Страница 18: ...16 ZUV User s manual CONTENTS Introduction CONTENTS MEMO...
Страница 44: ...2 18 Section 2 ZUV BASIC OPERATION ZUV User s Manual Section 2 Operating Modes MEMO...
Страница 68: ...3 24 Section 3 SETUP ZUV User s Manual Section 3 Setting the System Environment MEMO...
Страница 120: ...5 12 Section 5 APPENDIX ZUV User s Manual Section 5 Requirements from Regulations and Standards...
Страница 123: ...5 15 ZUV User s Manual Section 5 APPENDIX Section 5 INDEX MEMO...
Страница 125: ......