4-36
Section 4
CONNECTING TO EXTERNAL DEVICES
ZUV
User’s Manual
Section 4
Connection by USB/RS-232C
HDATACLR command
This command initializes the cumulative irradiation energy and power tuning data
saved on the controller.
After the head is replaced, be sure to initialize this data.
(1)
Parameter description
The following parameter can be set by the HDATACLR command:
(2)
Command/response examples
The following shows an example of this command:
The following shows an example of a response to this command.
HDATACLR <Channel No.>
OK
Parameter
Setting Range/Output Range
Channel No.
1: Channel 1
2: Channel 2
3: Channel 3
4: Channel 4
HDATACLR 1
←
The irradiation cumulative energy and
power tuning data for channel 1 is
initialized.
OK
←
Command execution ends successfully.
Command
Delimiter
Delimiter
Response
Delimiter
Delimiter
Delimiter
Delimiter
Delimiter
Delimiter