5-3
ZUV
User’s Manual
Sec
tion 5
APPENDIX
Section 5
Error Messages and Countermeasures
Error Messages and Countermeasures
Display Details
Cause
Countermeasure
SYSTEM ERROR
(CPU COMM)
The ZUV-C30H may have
malfunctioned.
Contact your OMRON representative.
SYSTEM ERROR
(ROM)
An error has occurred in the save
memory.
Clear flash memory in the controller by
holding down the SET key for 3 seconds.
Contact your OMRON representative if
normal operation is not restored.
SYSTEM ERROR
(HEAD)
The head has become loose or
malfunctioned, or the cable
disconnected during operation.
Check to see if the head is connected, and
turn ON the power again. Replace the
head or cable if normal operation is not
restored.
EMERGENCY HALT
The controller is in the emergency stop
state because the emergency stop input
is ON.
Set the emergency stop input to OFF.
• Open the emergency stop input if the
emergency stop input setting is [SHORT
(default)].
• Short-circuit the emergency stop input if
the emergency stop input setting is
[OPEN].
HEAD LONGEVITY
The cumulative alarm threshold
exceeds the cumulative irradiation time.
Replace with a new head.
CHANGE HARD
-WARE KEY TO
[LOCK]
The power supply was turned ON when
the key switch was in the "READY"
position.
Change the position of the key switch to
"LOCK".
CHANGE HARD
-WARE KEY TO
[READY]
The key switch is in the "LOCK" position
when changing to the READY mode.
Change the position of the key switch to
"READY".
MAX TIME ERROR
The total time exceeds 999.9(S) when
determining the irradiation time of the
irradiation pattern or when displaying
[VIEW].
Press [OK], and reset the total time to less
than 999.9(S).
UV LEVEL IS INVALID
The analog input value exceeds the
range of the received light amount, for
which the reference setting is available.
Check to see if the jig is connected, or
replace the head.
NOT TEACHING
Power tuning is executed with the
reference value unregistered.
Register the reference value, and then re-
execute power tuning.
UV LEVEL IS INVALID
The analog input value exceeds the
range of the received light amount, for
which power tuning is available.
Check to see if the jig is connected, or
replace the head.
ANALOG INPUT
ERROR
Teaching or power tuning cannot be
executed because the analog input
value is not stable.
The UV light incident to the UV luxmeter
may be unstable. Check the items below:
• Stabilize the installations of the UV
luxmeter and the UV irradiation head.
• Execute teaching/power tuning in low-
noise conditions.
Содержание ZUV Series
Страница 1: ...Smart Curing System User s Manual Cat No Z281 E1 02 ZUV Series Mega Power Type...
Страница 18: ...16 ZUV User s manual CONTENTS Introduction CONTENTS MEMO...
Страница 44: ...2 18 Section 2 ZUV BASIC OPERATION ZUV User s Manual Section 2 Operating Modes MEMO...
Страница 68: ...3 24 Section 3 SETUP ZUV User s Manual Section 3 Setting the System Environment MEMO...
Страница 120: ...5 12 Section 5 APPENDIX ZUV User s Manual Section 5 Requirements from Regulations and Standards...
Страница 123: ...5 15 ZUV User s Manual Section 5 APPENDIX Section 5 INDEX MEMO...