15
4.3.3.
Sensor Connection
Connect the sensor/s to the terminal block/s at the top of data logger:
Figure 11:
OM-DAQLINK logger sensor inputs
Sensors do not have to be added successively. You may only configure In-4, or configure In-1
and In-3, for example, when setting up the logger via the software.
4.3.4.
Programming Limitations for OM-DAQLINK-TEMP and OM-
DAQLINK-TEMPRH
The standard sampling rate for all sensors on all inputs is from a maximum of one sample every
second to a minimum of one sample every 18 hours. However, there are certain limitations
using a specific combination of sensors, which must be taken into account when programming
the OM-DAQLINK data loggers.
Note:
The software integrates all programming limitations automatically when configuring the
loggers.
Please refer to the table below for OM-DAQLINK programming limitations, where:
PT100, Thermocouple J, K or T
=
A
Internal Digital Temperature (on OM-
DAQLINK-TEMPRH)
=
B
Sensor
Combination
Maximum Sampling Rate
Example Setup
3 x A
4 x A
A + B
One sample every three seconds
In-1:
PT-100
In-2:
TC-J
In-3:
TC-J
(2 x A) + B
(3 x A) + B
One sample every four seconds
Internal:
Temp
In-1:
TC-T
In-2:
PT-100
(4 x A) + B
One sample every five seconds
Internal:
Temp
In-1:
TC-T
In-2:
TC-J
In-3:
PT-100
In-4:
PT-100
Table 1:
Data Logger programming limitations