60
Note:
This feature works only if there are no other commands in the queue following the
Download process.
Reset Alarm
When the logger is in a state of alarm, this option will reset the alarm state to normal on both the
logger and the software. The visual and audible alarms will return to normal. If after reset the
logger is still in alarm conditions, then the visible and audible alarms will return.
Calibration
Refer to section
5.17 for an overview of the Calibration feature.
Note:
A logger can’t be calibrated while it is in
Run
mode. The option will be grayed out. You
must
Stop
the logger first.
Cancel Firmware Update
When a firmware update is in progress, select this option to cancel the update and continue
using the previous firmware version.
Setup
Refer to section
5.15 for an overview of how to setup the logger.
Stop
Sends a command to the logger to stop logging data.
Run
Sends a command to the logger to start logging data. This command will also clear the existing
logger memory.
Turn Off
This option will turn the unit off and is password enabled. As a precaution to avoid any data loss
all data not yet transmitted will first be downloaded to the computer prior to turning the unit off.
5.15.
Configuring the Logger
DataSuite software allows you to configure each OM-DAQLINK logger, via direct USB
connection.
To configure the logger once it’s detected by the DataSuite software, right-click the Logger icon
in the Map View and select
Setup
. The Setup window is described below.
5.15.1.
OM-DAQLINK Setup Window – Device Setup Tab
The OM-DAQLINK Setup window has two tabs,
Device Setup
and
Alarm Setup
.