92
A.2.
OM-DAQLINK-TEMP and OM-DAQLINK-TEMPRH
Outputs
Alarm Output (Output 1)
External Power Excitation
(transducers usage)
Open collector
Close position resistance:
50 Ù
Max. Load:
50 mA, 3 V DC
12 VDC
@
2 A
Reset fuse overload protection
A.3.
Logger Input Specifications OM-DAQLINK-TEMP
and OM-DAQLINK-TEMPRH
Internal PT-100
Temperature
(OM-DAQLINK-TEMP)
Internal Digital
Temperature
(OM-DAQLINK-TEMPRH)
Internal Digital
Relative Humidity
(OM-DAQLINK-TEMPRH)
Type:
PT-100
Range:
-
20 to 50 ˚C
Resolution:
0.1 ˚C
Accuracy:
±0.3 ˚C
Type:
Digital sensor
Range:
-
20 to 50 ˚C
Resolution:
0.1 ˚C
Accuracy:
±0.5 ˚C
Range:
5 to 95 %
Resolution:
0.5%
Accuracy:
3% (DataSuite software)
4% (Logger LCD)
Temperature
Thermocouple J
Temperature
Thermocouple K
Temperature
Thermocouple T
Range:
-
200 to 1,000 ˚C
Resolution:
0.1 ˚C
Accuracy:
-200 to
– 60 ˚C ±0.5 %
-
60 to 60 ˚C ±0.5 ˚C
60 to 1,000 ˚C ±0.5 %
Cold junction
compensation:
±0.3 ˚C
Range:
-
200 to 1,000 ˚C
Resolution:
0.1 ˚C
Accuracy:
-200 to
– 60 ˚C ±0.5 %
-
60 to 60 ˚C ±0.5 ˚C
60 to 1,000 ˚C ±0.5 %
Cold junction
compensation:
±0.3 ˚C
Range:
-
200 to 400 ˚C
Resolution:
0.1 ˚C
Accuracy:
-200 to
– 60 ˚C ±0.5 %
-
60 to 60 ˚C ±0.5 ˚C
60 to 400 ˚C ±0.5 %
Cold junction
compensation:
±0.3 ˚C