7
7.1
Troubleshooting Guide
67
Remote control from the
connected equipment is not
available.
The remote cable is connected improperly.
Connect it properly as described in “Chapter 2
Installation and Connection” in “INSTRUCTIONS
(BASIC)” or “Chapter 2 Installation and Connection” on
page 15 in “INSTRUCTIONS (ADVANCED).”
The remote cable is broken.
Replace the new remote cable.
The remote cable is subjected to an excessive
force.
Change the cable layout to remove the excessive force.
The endoscope or scope cable is not connected. Connect the endoscope or scope cable as described in the
instruction manual for the connected equipment.
The connected equipment is turned OFF.
Turn the connected equipment ON as described in its
instruction manual.
This instrument is turned OFF.
Turn this instrument ON.
The connected equipment is set improperly.
Set the connected equipment properly as described in
“3.1 Setup of Connected Equipment” on page 25, or in
“3.3 Setup of Connected Equipment” in
“INSTRUCTIONS (BASIC).”
The function of the REMOTE 2 terminal is set
improperly.
Set it properly as described in “Setting “RM2 FUNC”” on
page 61 and in the instruction manual of the connected
equipment.
The communication speed (BAUDRATE) is set
improperly (for RS-232C communication).
Set it properly as described in “Setting “BAUDRATE””
on page 59 and in the instruction manual of the connected
equipment.
A communication problem has occurred
between this instrument and the connected
equipment.
Turn this instrument OFF, then ON again.
Turn the connected equipment OFF, then ON again.
The CVP counter is not
displayed.
The remote cable is connected improperly.
Connect it properly as described in “Chapter 2
Installation and Connection” in “INSTRUCTIONS
(BASIC)” or “Chapter 2 Installation and Connection” on
page 15 in “INSTRUCTIONS (ADVANCED).”
The remote cable is broken.
Replace the new remote cable.
The remote cable is subjected to an excessive
force.
Change the cable layout to remove the excessive force.
The connected equipment is turned OFF.
Turn the connected equipment ON as described in the
instruction manual.
This instrument is turned OFF.
Turn this instrument ON.
The connected equipment is set improperly.
Set it properly as described in “3.3 Setup of the
Connected Equipment” in “INSTRUCTIONS (BASIC)”
or “3.1 Setup of Connected Equipment” on page 25 in
“INSTRUCTIONS (ADVANCED).”
The communication rate (BAUDRATE) is set
improperly.
Set it properly as described in the description on “Setting
“BAUDRATE”” on page 59 as well as the instruction
manual for the connected equipment.
A trouble occurred in the communication with
the connected equipment.
Turn this instrument OFF then ON again.
Turn the connected equipment OFF then ON again.
The video system center is remove data mode.
Remove data mode is cancelled with reference to the
instructions manual of the video system center.
Irregularity
Possible cause
Solution