![Olympus IPLEX NX IV9000N Скачать руководство пользователя страница 104](http://html.mh-extra.com/html/olympus/iplex-nx-iv9000n/iplex-nx-iv9000n_instructions-manual_735024104.webp)
96
Message
Cause and recommended action
<NO RECORDING MEDIA.>
The SDHC card or the USB flash drive is not loaded.
→
Load the SDHC card or the USB flash drive you use and try
again. Or turn OFF the instrument and ON again.
<RECORDING MEDIA FULL.>
The SDHC card or the USB flash drive is full.
→
Delete the unnecessary data to free up space, or replace
with a new SDHC card or a new USB flash drive.
<THIS IMAGE CANNOT BE RETRIEVED.>
The image is not recorded with this instrument.
→
Only images recorded with this instrument can be
displayed.
<RECORDING MEDIA ERROR.
PLEASE FORMAT THE RECORDING MEDIA.>
The format of the SDHC card or the USB flash drive cannot
be recognized.
→
Format the SDHC card or the SB flash drive with this
instrument.
<THE SPECIFIED FILE NAME ALREADY
EXISTS.
PLEASE CHOOSE ANOTHER NAME.>
The file name exists.
→
Specify the file name other than the existing file name.
<THE SPECIFIED FOLDER NAME ALREADY
EXISTS.
PLEASE CHOOSE ANOTHER NAME.>
The folder name exists.
→
Specify the folder name other than the existing folder name.
<RECORDING MEDIA IS LOCKED. PLEASE
UNLOCK.>
The SDHC card is locked.
→
Unlock the SDHC card you will use and insert the SDHC
card in this instrument.
<INCOMPARABLE IMAGE RESOLUTION.>
The resolution of the live image is different from that of the
played back image.
→
Specify the file with the same resolution as the live image.
<The image cannot be displayed in 3D view.
Acquire the stereo image again.>
The image is not appropriate for 3D view.
→
Acquire the stereo image again.
<OVER CURRENT (USB).
PLEASE TURN OFF THE POWER.>
The recommended USB flash drive is not used.
→
Stop inspection, and turn OFF the instrument and ON again.
Use the recommended USB flash drive.
A USB device other than the USB flash drive is connected.
→
Disconnect all USB devices other than the USB flash drive.
88
Troubleshooting
Inspect the instrument as described in “3 Pre-operation preparation and inspection” (page 22).
If there are any obvious malfunctions, do not use the instrument and contact Olympus for
repairs. If a slightest irregularity is suspected, do not use the instrument and perform the actions
described in “8-1 Troubleshooting guide”. If the problem cannot be resolved by taking the
remedial action described, stop using the instrument and contact Olympus for repair.
8-1 Troubleshooting guide
1
Error messages
Содержание IPLEX NX IV9000N
Страница 8: ......
Страница 19: ...11 IPLEX NX Scope unit optional Model name serial number Label without description Caution Conformity label ...
Страница 121: ...113 IPLEX NX MEMO ...
Страница 124: ...MEMO ...