IPLEX GAir
6.1 Scaler measurement function
6.1.1 Scaler measurement
Scaler measurement measures the length of an object using the reference length value that is
preset according to a known length inside the observation image.
NOTE:
Therefore, if the reference length is incorrect, the correct measurement accuracy cannot be
obtained. Note that the measurement is made based on the assumption that the reference object
and the object to be measured are on the same plane vertical to the optical axis.
6.1.2 Scaler measurement screen
Explanation of the scaler measurement screen
No.
Name
Function
1
Message box
Displays information related to operation.
2
Measurable region
Displays the measurable region with solid line.
3
Recommended measurement
region
Displays the recommended measurement region with
dotted line.
4
Reference length
Inputs the dimensions of the parts of which length is
known.
5
[CURSOR/OK] button
Operates the cursor and inputs the reference length on
the screen.
6
Point
Reference point or measuring point that has been input
7
Cursor
Specifies the reference point or measuring point.
8
Measurement value
Displays the measurement result of the input points.
6.
Measurement functions
119