![Olympus IPLEX GAir IV9000GA Скачать руководство пользователя страница 116](http://html.mh-extra.com/html/olympus/iplex-gair-iv9000ga/iplex-gair-iv9000ga_instructions-manual_735023116.webp)
Main
menu
Sub
menu
Item
You can display the software version of the base unit, initial
data version and the upgrade information of the system.
<START UP>
Saves the settings when the power was turned OFF.
By saving the settings when the power was turned OFF, you
can specify whether to automatically apply such settings or
not when the power is turned ON next time.
l
<OFF>: Disable. (Applies the default settings when
the power is turned ON next time.)
l
<ON>: Saves the settings when the power was
turned OFF, and applies such settings when
the power is turned ON next time.
(Settings that can be saved: Brightness
adjustment, zoom factor, gain mode, and high
beam or normal beam (eco))
<LCD
BACKLIGHT>
Specifies the brightness of the back light of the LCD
monitor.
The brightness level increases from bottom to top (5 levels).
<ANGLE SPEED>
Specifies the angulation speed.
The angulation speed can be selected between <SLOW> and
<FAST> (2 levels).
<LENGTH UNIT>
Selects the unit of insertion length.
You can select <meter> or <foot/inch>.
<LENGTH
FORM>
Selects the method to display the insertion length.
You can select <SECTION LENGTH> or<COMBINED LENGTH>.
See "Displaying the insertion length" (page 99).
<CLEAR DRAIN
ERROR>
Clears the drain error of the compressor.
This option cannot be used if there is no drain error when you
turn ON the power of this instrument or after the drain error
is cleared.
See "Removing water from the drain" (page 133).
<REMOTE
CONTROLLER
VIBRATION>
Sets the vibration of the remote control unit.
You can set whether to vibrate the remote control unit or not
110