![OLS 2800000 Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/ols/2800000/2800000_operators-manual_732074058.webp)
CERO_Operators Guide_Version 2018-12_01
58
8. TROUBLESHOOTING
8.1 Temperature
8.1.1 Temperature Too High
Possible explanation
Recommended action
1. Offset value too high
Readjust temperature offset (chapter 5.4)
2. Convection Channel defect
Replace convection channel (chapter 7.4)
3. Room temperature too high
Verify that room temperature is at least 7 °C below set
temperature
4. Electronics fan defect (at back
of the CERO)
Call customer support
5. Temperature sensor defect
Call customer support
8.1.2 Temperature Too Low
Possible explanation
Recommended action
1. Offset value too low
Readjust temperature offset (chapter 5.4)
2. CERO covers open
Close the CERO covers
3. Temperature sensor defect
Call customer support
8.1.3 Temperature Unstable
Possible explanation
Recommended action
1. Convection channel defect
Replace convection channel (chapter 7.4)
2. CERO covers open
Close the CERO covers
3. Room temperature fluctuating
Place the CERO in another location
4. Temperature sensor defect
Call customer support
5. Electrical noise
Remove nearby sources of RFI, including motors, arcing
relays or radio transmitters