5.4 Set Temperature Offset Compensation
The Temperature Offset Compensation may be used to adjust the temperature to your specific
standards, especially before the first operation of the instrument.
1.
Place a CEROtube with 40 mL pre-warmed (37°) water in the position 2 of the CERO (see
Chapter 5.7).
2.
Start the temperature regulation at 37°C (as described in Chapter 5.2).
3.
After 1 hour, open the Chamber Access Port on the Sliding Top Cover by removing the Coin
Screw.
4.
Measure the temperature of the liquid in the tube, using a standard laboratory temperature
probe.
5.
Go to the System Menu and select Offset Calibration (Figure 18).
Figure 18: Ca lib ra tion Offs et Window
6.
Select Temperature Offset and rotate the click-wheel to set the Temperature Offset.
NOTE
The Temperature Offset is defined as follows:
Offset = temperature of extern probe – temperature of CERO