CERO_Operators Guide_Version 2018-12_01
48
6.4.4 Adjust the Settings for the Color Monitoring
1. In the System Menu, go to Settings.
2. Enter the Reader Window and select pH Settings (figure 35).
3. Verify that the Color Mode is enabled and enable it if necessary (figure 36).
4. Adjust the desired parameters:
- Rotation Speed: speed of the tube rotation during measurement
- Settling Time: time allowed for the cells to settle before the measurement starts
6.4.5 Adjust the Settings for the pH Monitoring
1. In the System Menu, go to Settings.
2. Enter the Reader Window and select pH Settings (figure 35).
5. Verify that the pH Mode is enabled and enable it if necessary (figure 37).
Figure 37: pH Settings (pH Monitoring Mode)
6. Adjust the desired parameters:
- Rotation Speed: speed of the tube rotation during measurement
- Settling Time: time allowed for cell settlement before the measurements starts
- pKa Default: pKa of phenol red at bei 37 °C
- εΤ: extinction coefficient of the phenol red
- Ci Default: final concentration of phenol red in the medium used and accounting for all
additional dilutions