CERO_Operators Guide_Version 2018-12_01
57
7.7.2 Decontamination pH-Reader
ATTENTION
Do not use disinfecting materials which contain hypochlorite (Javelle water,
Chlorox) or bleaching fluids.
1. Make sure that the CERO is switched off before performing the decontamination procedure.
2. Perform the weekly maintenance (chapter 7.7.1).
7.8 Check the Memory Space Available
1.
Go to the System Menu and select Information (figure 49)
2.
Check the amount of memory space available for saving new Outputs under Storage Data.
3.
We recommend freeing memory space and export or delete Outputs if Storage Data is 80%
full.
Figure 49: System Informa tion Window
7.9 Yearly Maintenance
We recommend that you contact your local service and support team for the yearly
maintenance (see contact information at the back of this Operators Guide).
NOTE
The CERO provides enough memory space for saving about 5000 standard
experiment Outputs.
If there is not enough memory space older experiments outputs will be
automatically deleted.