Ranger™ Count 3000 Series EN-13
4.2 Cal Menu
Enter this menu to perform calibrations.
Initial Calibration
When the scale is operated for the first time, a span calibration is recommended to ensure
accurate weighing results. Before performing the calibration, be sure to have the appropriate
calibration weights as listed in table 4-1. Ensure that the LFT switch/calibration lock is set to
unlocked position. See figure 5-1.
Or adjust the GEO setting according to your location (see table 4-2).
Procedure:
Press and hold
Menu
until [
mMeNU
] (Menu) is displayed. When the button is released, the display
will show [
C.A.L
]. Press
Yes
to accept. [
SpaN
] will then be shown. Press
Yes
to begin the span
calibration. [
0
kg
] will be displayed. Press
Yes
to accept. [
--C--
] will be displayed while zero
reading is stored. Next, the display shows the calibration weight value. Place the specified
calibration mass on the pan. Press
Yes
to accept the weight or
No
to select an alternate weight.
[
--C--
] will be displayed while the reading is stored. The display will show [
done
] if the calibration
was successful. The scale returns to the previous application mode and is ready for use.
TABLE 4-1
Required Span Calibration Mass (sold separately)
Capacity Mass
*
Capacity Mass
*
1500g
1.5kg / 3lb
15000g
15kg / 30lb
3000g
3kg / 6lb
30000g
30kg / 60lb
6000g
6kg / 15lb
Note:
When active unit is g or kg, the calibrating unit will be in kg.
When active unit is lb, oz or lb:oz, the calibrating unit will be in lb.
For linearity calibration, the calibration Mass is fixed.
Mid-point is always half of full capacity.