Ranger™ Count 3000 Series EN-17
4.4 Readout Menu
Enter this menu to set user preferences. Default settings are in
bold
.
Reset
[
reset
]
Reset the Read menu to factory defaults.
NO
= not reset
YES
= reset
Stable Range
[
Stable
]
Set the amount the reading can vary while the stability symbol remains on.
0.5d
= 0.5 scale division
1d
= 1 scale division
2d
= 2 scale division
5d
= 5 scale division
Filter
[
FILtEr
]
Set the amount of signal filtering.
LOW
= less stability, faster stabilization time
MED
= normal stability, stabilization time
HI
= greater stability, slower stabilization time
AZT
[
AZt
]
Set the automatic zero tracking functionality.
OFF
= disabled
0.5d
= the display will maintain zero until a change of 0.5 divisions per second has
been exceeded.
1d
= the display will maintain zero until a change of 1 divisions per second has been
exceeded.
3d
= the display will maintain zero until a change of 3 divisions per second has been
exceeded.
Light
[
LIGHt
]
Sets backlight functionality.
OFF
= always off
ON
= always on
AUTO = turns on when a button is pressed or the displayed weight changes.
Auto off
[
A.OFF
]
Set the automatic shut off functionality.
OFF
= disabled
1
= powers off after 1 minute of no activity
5
= powers off after 5 minute of no activity
10
= powers off after 10 minute of no activity
End Readout
[
End
]
Advance to the next menu or return to the top of the current menu.
Reset:
no
, yes
Stable Range:
0.5,
1
, 2, 5
Filter:
low,
medium
, high
Auto-Zero Tracking: off,
0.5
, 1, 3
Light:
off, on,
auto
Auto off:
off
, 1, 5, 10
End Readout:
Exit menu