CHAPTER 3 - MAINTENANCE / REPAIR PROCEDURES
Ohaus Corporation www.ohaus.com
40
Explorer
®
Series Service Manual
3. Remove the
Connect PCBA
•
Disconnect the Connect Cable to the Main PCBA.
•
Remove the 2 retaining screws, the PCB is now free
•
During disassembly and re-assembly note how the switch engages the Switch
Actuator.
4. Remove
Main PCB
Note:
For EX12001, EX24001 and EX35001 models please see section 5.
•
Remove the left IR Sensor PCBA (see above).
•
Disconnect the cables along the top edge of the PCB.
•
Loosen the 2 retaining screws approximately 2 turns
•
Rotate the retaining washers so that it no longer engages the PCB, slide out the PCB
•
Disconnect the Power Jack and the Ribbon Cable to the Connector PCBA cables
Top edge connectors
Real Time Clock battery
Содержание Explorer Balances
Страница 1: ...Explorer Balances Service Manual...
Страница 2: ......
Страница 6: ......
Страница 82: ......