CHAPTER 3 - MAINTENANCE / REPAIR PROCEDURES
Ohaus Corporation www.ohaus.com
28
Explorer
®
Series Service Manual
3.2.5 Opening the Base Module (EX12001, EX24001, EX35001)
1. Prepare the balance as instructed above.
2.
Remove the 2 screws that hold the Spider to the Top Housing and lift off the Spider.
3.
Remove the 7 screws that hold the Top Housing to the Base.